    Product Specifications
    Assay TypeSandwich-ELISA
    AnalyteTPO
    Format96T(8×12 strips)
    ReactivityHuman
    Regulatory StatusRUO
    Sensitivity<12.5 pg/mL
    Standard Curve Range12.5 pg/mL-400 pg/mL
    Assay Time3 hr 20 min
    Suitable Sample TypeFor the detection and quantitative determination of TPO in human serum and cell culture supernates.
    Sample volume100 uL
  • Product Overview
    resDetect™ Human TPO ELISA Kit (Residue Testing) is based on the ELISA sandwich method and is used to detect and quantitatively determine GMP Human TPO residues in cell culture supernatants, serum, and plasma. The kit contains GMP human TPO (ACROBiosystems, cat# GMP-THNH25) to ensure accurate assay results and is designed to provide a reliable solution for CAR-T product quality assessment during drug development and CMC quality control stages. It can also be used as a universal detection tool for the quantitative determination of Human TPO.
  • Reconstitution
    Please see Certificate of Analysis for details of reconstitution instruction and specific concentration.
  • Storage
    1. Unopened kit should be stored at 2℃-8℃ upon receiving.

    2. Find the expiration date on the outside packaging and do not use reagents past their expiration date.

    3. The opened kit should be stored per components table. The shelf life is 30 days from the date of opening.

    Materials Provided
    IDComponentsSize
    RES034-C01Pre-coated Anti-TPO Antibody Microplate1 plate(8×12 strips)
    RES034-C02Human TPO Standard20 μg
    RES034-C03Biotin-Anti-TPO Antibody20 μg
    RES034-C04Streptavidin-HRP50 μL
    RES034-C0510×Washing Buffer50 mL
    RES034-C062×Dilution Buffer50 mL
    RES034-C07Substrate Solution12 mL
    RES034-C08Stop Solution7 mL
Typical Data Please refer to DS document for the assay protocol.
 TPO TYPICAL DATA

For each experiment, a standard curve needs to be set for each micro-plate, and the specific OD value may vary depending on different laboratories, testers, or equipments. The following example data is for reference only.

Validation
Dilution Linearity

To assess the linearity of the assay, samples spiked with high concentrations of TPO were serially diluted with calibrator diluent to produce samples with values within the dynamic range of the assay.

 TPO DILUTION LINEARITY
Intra-Assay Statistics

Three samples of known concentration were tested ten times on one plate to assess intra-assay precision.

 TPO INTRA-ASSAY STATISTICS
Inter-Assay Statistics

Three samples of known concentration were tested in three separate assays to assess inter-assay precision.

 TPO INTER-ASSAY STATISTICS
Recovery

Five parts of blank T cell culture supernatant were added with different concentrations of human TPO, and the T cell culture supernatant without human TPO was used as background to calculate the recovery rate. The range of the recovery rate is 94.7-108.9%, and the average recovery is 102.1%.

 TPO RECOVERY
  • Clinical and Translational Updates

Comments (0)


ETA of in-stock products: 2 business days

Price(USD) : $360.00

Promotion & Exhibitions



Datasheet & Documentation

Request for DMF
DMF (Drug Master File)

Related Products

New Product Launch

Questions & Comments

This web search service is supported by Google Inc.

totop
Call us
Call us
North America:
+1 800-810-0816 (Toll Free)
Asia & Pacific:
+86 400-682-2521
Fax
Fax
+1 888-377-6111
Address
Address
1 Innovation Way, Newark, DE 19711, USA

Leave a message